Patient Spirometry
There are three keys on the compact airway modules E-CAiOV, E-CAiOVX, E-COV,
E-COVX:
Save Loop
Saves the currently active loop with corresponding numeric
data.
Print Loop
Prints the current pressure-volume (Paw-Vol) or flow-volume
(Flow-Vol) loop with corresponding numeric data.
Change Loop
Toggles between a Paw-Vol and a Flow-Vol loop.
Preparing the Patient Spirometry measurement
1.
Take a new Patient Spirometry tube and connect the tube to the D-lite(+)/Pedi-lite(+)
sensor by inserting the angle connectors in the sensor connectors. Place all
D-lite(+)/Pedi-lite(+) ports upwards with approximately a 45 °C tilt to prevent
condensed water from entering the sensor interior and the tubings.
2.
Connect the other end of the Patient Spirometry tube to the pressure connectors
on the module.
3.
Connect a gas sampling line to the Luer connector on the other side of the
D-lite(+)/Pedi-lite(+) sensor.
4.
Connect the other end of the gas sampling line to the sampling line connector on
the module’s water trap.
5.
Make sure that the connections are tight.
6.
Select the correct sensor type.
7.
Connect the D-lite(+)/Pedi-lite(+) between the Y-piece and the intubation tube
in the breathing circuit.
Checking the Patient Spirometry measurement
1.
Check that the water trap is empty.
2.
Occlude the sampling line and check that the
Sample line blocked
message
appears within 30 seconds and gas waveforms are showing zero at the same time.
3.
Check that the loops are whole. A gap between the starting and ending points
may indicate a leak.
Using the Patient Spirometry measurement
Selecting the Patient Spirometry sensor type
When monitoring pediatric patients with tidal volumes of 15 ml to 300 ml, use the
Pedi-lite/Pedi-lite+ sensor. For other patients, use the D-lite/D-lite+ sensor. Select the
sensor type accordingly.
1.
Select the Patient Spirometry parameter window.
2.
Select
Setup
.
3.
Select the sensor type (
Pediatric
or
Adult
) from the
Sensor Type
list.
2062971-001
CARESCAPE Modular Monitors
325
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...