ECG
Setting HR alarm limits for a single HR source
1.
Select the HR parameter window.
2.
Select the
Alarms
tab.
3.
Check that the required alarm,
HR
,
PVC
, or
SVC
, is turned on.
If a feature is not active, the alarm limits are greyed out.
4.
Select
Alarm On
to set the alarms.
5.
Adjust the alarm limits with the arrows.
Setting HR/PR alarm limits for multiple HR sources
1.
Select the HR parameter window.
2.
Select the
HR/PR Alarms
tab.
3.
Check that the required alarm is turned on.
If a feature is not active, the alarm limits are greyed out and the
Alarm On
check
box is not selected.
4.
Select the
Alarm On
check box for those alarms you wish to set.
5.
Adjust the alarm limits with the arrows.
Setting PVC alarm limits
Available with the full arrhythmia license only.
1.
Select the HR parameter window.
2.
Select the
PVC/SVC Alarms
or
PVC Alarms
(TRAM or telemetry transmitters) tab. If
the heart rate setting is
Single
, select the
Alarms
tab.
3.
Check that the
PVC
alarm is turned on.
If a feature is not active, the alarm limits are greyed out.
4.
Select
Alarm On
to set the alarms.
5.
Adjust the alarm limits with the arrows.
Setting SVC alarm limits
Available with the full arrhythmia license only. Not available with combination
monitoring or TRAM.
1.
Select the HR parameter window.
2.
Select the
PVC/SVC Alarms
tab. If the heart rate setting is
Single
, select the
Alarms
tab.
3.
Check that the
SVC
alarm is turned on.
If a feature is not active, the alarm limits are greyed out.
4.
Select
Alarm On
to set the alarms.
5.
Adjust the alarm limits with the arrows.
Selecting the HR alarm range
With the
Single
heart rate setting only.
164
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...