Messages
Message
Location
Explanation
What to do
●
Alarm volume
changed
●
al. area
The network connection is lost,
and the local alarm volume is
increased.
●
Readjust volume if desired.
●
Alarms acknowl-
edged remotely
●
al. area
The alarms have been
acknowledged from the central
station.
●
No action required but you
may wish to confirm in alarm
history which alarm has been
acknowledged.
●
Alarms audio paused
remotely
●
al. area
The alarms were remotely
paused from the central station.
●
You can activate the alarms
by pressing the audio pause
key.
●
Alarms audio paused
from telemetry
●
al. area
The alarms were remotely
paused from the telemetry
transmitter.
●
You can activate the alarms
by pressing the audio pause
key.
●
All monitors
disconnected
●
al. area
The monitor is disconnected
from the network.
●
Re-establish connection.
●
If the problem persists,
contact authorized service
personnel.
●
Application
error: iPanel
●
al. area
Connection to iPanel was lost.
●
Re-establish the connection.
●
If the problem persists,
contact authorized service
personnel.
●
Application error: pdf
●
al. area
PDF viewer closes unexpectedly.
●
Try to open the MUSE 12SL
report again.
●
If the problem persists,
contact authorized service
personnel.
●
Application
error: Webmin
●
al. area
Webmin application closes
unexpectedly.
●
Try to open Webmin again.
●
If the problem persists,
contact authorized service
personnel.
●
Audio set OFF
remotely
●
al. area
The alarms have been turned off
from the central station.
●
You can change the selection
through
Alarm Setup
>
Audible & Visual
.
●
Barcode scanned
●
al. area
All data has been successfully
stored to the monitor.
●
No action required.
●
Barcode too long
●
al. area
The maximum length of the
barcode has been exceeded.
●
Verify the information read by
the barcode reader and edit if
necessary.
●
Battery A failure
●
Battery B failure
●
al. area
Battery A or B is faulty.
●
Replace the battery if
necessary.
●
Battery failure
●
al. area
The monitor battery is faulty.
●
Replace the battery if
necessary.
506
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...