Messages
Message
Location
Explanation
What to do
●
P1 sys high
/
P1 sys
low
●
P1 mean high
/
P1 mean low
●
P1 dia high
/
P1 dia
low
to
●
P8 sys high
/
P8 sys
low
●
P8 mean high
/
P8 mean low
●
P8 dia high
/
P8 dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
P1 zeroing failed
to
●
P8 zeroing failed
●
param.
The channel has not been zeroed
successfully.
●
Repeat the zeroing.
●
P7 connected
●
al. area
The channel has been
connected.
●
No action required.
●
P8 connected
●
al. area
The channel has been
connected.
●
No action required.
●
PA sys high
/
PA sys
low
●
PA mean high
/
PA mean low
●
PA dia high
/
PA dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
PR(ABP) high
/
PR(ABP)
low
●
PR(Art) high
/
PR(Art)
low
●
PR(Fem) high
/
PR(Fem) low
●
PR(UAC) high
/
PR(UAC)
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Pressure
measurement
removed
●
al. area
The acquisition module has been
removed.
●
Reconnect if necessary.
●
Pressure Sensed
●
param.
Pressure has been sensed during
zeroing.
●
Open the venting stopcock to
air.
●
RAP sys high
/
RAP sys low
●
RAP mean high
/
RAP mean low
●
RAP dia high
/
RAP dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
2062971-001
CARESCAPE Modular Monitors
479
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...