Messages
Message
Location
Explanation
What to do
●
RVP sys high
/
RVP sys
low
●
RVP mean high
/
RVP mean low
●
RVP dia high
/
RVP dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Sensor P1 failed
to
Sensor P8 failed
●
al. area
IP: Faulty or disconnected
sensor.
●
Check the cable and
connections.
●
Replace the transducer.
●
Sensor
●
param.
IP: Faulty or disconnected
sensor.
●
Check the cable and
connections.
●
Replace the transducer.
●
UAC disconnect
●
al. area
Invasive pressure line is
disconnected.
●
Check the patient status.
●
Check connections.
●
If pressure drops because of
zeroing, perform the zeroing
process.
●
UAC sys high
/
UAC sys low
●
UAC mean high
/
UAC mean low
●
UAC dia high
/
UAC dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
UVC sys high
/
UVC sys low
●
UVC mean high
/
UVC mean low
●
UVC dia high
/
UVC dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Zero adj >100 mmHg
●
param.
IP channel zeroed to over 100
mmHg pressure.
●
Repeat the transducer
zeroing.
●
Replace the sensor.
●
Replace the transducer.
●
Replace the module.
●
If the problem persists,
contact authorized service
personnel.
●
Zeroed
●
param.
Zeroing was successful.
●
No action required.
Message is automatically
removed after 10 seconds.
480
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...