Pulse oximetry
WARNING
— FAILURE TO DETECT LETHAL ARRHYTHMIA — Always monitor
ECG for arrhythmia detection purposes. HR calculated from
pulsatile SpO
2
waveform may differ significantly from ECG HR
measured values. Users should be aware that the
SpO2 probe
off
and
No SpO2 pulse
technical alarms escalate no higher
than a
Medium
priority.
WARNING
— FAILURE TO DETECT LETHAL ARRHYTHMIA — The SpO
2
parameter pulsatile heart rate measurement is based on the
optical detection of a peripheral flow pulse and therefore may
not detect certain arrhythmias. The pulse oximetry parameter
should not be used as a replacement or substitute for ECG
based arrhythmia analysis.
WARNING
Using the
Maximum
sensitivity setting delays the
SpO2 probe
off
detection alarm.
WARNING
With deactivated
SpO2 probe off
alarm, keep the patient
under close surveillance.
WARNING
— MISSED ALARM — Check the SpO
2
measurement when
switching the SpO
2
measurement sources to avoid missed
SpO
2
alarms.
SpO
2
cautions
CAUTION
A damaged sensor or a sensor soaked in liquid may cause
burns during electrosurgery.
CAUTION
Prolonged monitoring or incorrect sensor application can
cause skin irritation or impaired circulation. It is recommended
that you check the probe site every four hours (more
frequently in case of poor perfusion or neonatal patients).
Refer to the instructions supplied with the sensor.
SpO
2
measurement limitations
●
Other E-modules than E-MASIMO and E-NSATX used for this measurement are not
suitable for use with neonatal patients.
●
The pulse oximeter cannot distinguish between oxyhemoglobin and
dyshemoglobins.
●
Poor perfusion may affect the accuracy of measurement, especially when using
an ear sensor.
●
To avoid erroneous measurements, do not use a blood pressure cuff on the same
limb as the SpO
2
sensor.
SpO
2
points to note
●
GE sensors are not made with natural rubber latex.
●
Use dry and clean sensors only.
●
Do not use damaged sensors.
●
Check that you are not re-using a disposable sensor or other disposable
accessories.
204
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...