WARNING
— PHYSICAL INJURY — Do not install the PDM above a patient.
Make sure the battery is completely inserted and the battery
door is completely closed. Falling batteries could seriously or
fatally injure neonatal or other vulnerable patients.
WARNING
— PHYSICAL INJURY — Do not install the PDM above a
patient. Leaks from the battery cells can occur under extreme
conditions. The liquid is caustic to the eyes and skin. If the
liquid comes in contact with eyes or skin, flush with clean
water and seek medical attention.
WARNING
— EXPLOSION OR FIRE — Using non-recommended batteries
could result in injury/burns to the patients or users. Only use
batteries recommended or manufactured by GE. The warranty
can be voided if non-recommended batteries are used.
TRAM module front view
TRAM and Tram-Rac modules with the B850 only.
The following is an example of a TRAM module (TRAM 451N5). The front panel
connectors are labeled. Some TRAM modules have different NIBP and SpO
2
connectors, but they are located in the same positions as those shown in the picture.
TRAM module invasive pressure connectors are labeled BP and non-invasive blood
pressure connectors are labeled NBP.
1. ECG (imp.resp.)
2. Temp/C.O.
3. BP1
4. BP2
5. BP3/4
6. NBP
7. SpO
2
8. Defib/Sync
9. Transport display
10. Power indicator
Single-parameter Tram-Rac front view
TRAM and Tram-Rac modules with the B850 only.
Single-parameter Tram-Rac modules collect the patient’s physiological data and send
it to the monitor for display. Single-parameter BP modules (in the picture) can only be
placed in the bottom two slots (slot 3 and slot 4) of a Tram-Rac 4A housing.
72
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...