Starting and ending monitoring
If a PDM or TRAM is connected to the monitor when the time is adjusted, the monitor
sends the new time to the module.
About case reset/patient discharge
Resetting a case/discharging a patient deletes all patient information from an
attached PDM or TRAM. If this is not desired, disconnect the PDM or TRAM from the
monitor before resetting a case/discharging the patient.
The monitor may be configured with an automatic case reset/patient discharge
timer. If this is configured and vital signs are no longer detected, monitoring will end
automatically after the configured time has elapsed.
The patient can be discharged remotely using a central station provided that this
option has been enabled. This option is not available in OR and PACU software
packages. However, if there is a pending automatic software or settings update that
will take place after the patient discharge, remote discharge is not possible. The
discharge option will then be disabled at the central station.
Pending software or setting activation
If there is a pending software or setting activation, it is indicated with the messages
Software activation after next discharge
/
Software activation after next case end
,
or
Setting activation after next discharge
/
Setting activation after next case end
.
When you reset a case/discharge a patient, this activation will automatically take
place. Follow the instructional texts that appear on screen.
NOTE
If the pending activation cannot be allowed to take place
and must be canceled for some reason, contact authorized
service personnel.
Residual physiological data
To ensure that no physiological data remains in the acquisition module or in the
bedside monitor after resetting a case/discharging the patient, do the following:
●
PDM: Do not disconnect the acquisition module from the bedside monitor before
ending a case or discharging a patient. You must also disconnect all the patient
cables from the patient. The PDM can continue to measure patient data with
battery power even when the module is not connected to the patient monitor.
●
Bedside monitor: Remove the acquisition modules from the monitor or disconnect
all the patient cables from the patient.
Resetting a case/discharging a patient
1.
Disconnect patient cables.
2.
Print necessary data and wait until the printing is completed.
3.
Select the patient information field on the screen.
4.
Select the
Patient
tab.
5.
Select
Reset Case
or
Discharge Patient
.
Monitor settings, including alarm limits, return to their default settings. All patient data
and trend data is removed from both the monitor and a connected PDM or TRAM.
2062971-001
CARESCAPE Modular Monitors
125
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...