Invasive pressures
Zeroing the invasive pressure transducers
1.
Level the transducer following your care unit’s policy (usually level of the
phlebostatic axis).
2.
Close the transducer stopcock to the patient and open the venting stopcock to air.
3.
If the pressure line you are trying to zero does not have the transducer open to air,
the message
Pressure Sensed
displays.
4.
You can zero all connected pressure transducers simultaneously by selecting
Zero
All Pressures
from the monitor’s main menu or from the remote control, or you
can zero a single active pressure transducer by selecting the invasive pressure
parameter window >
Setup
>
Zero
.
NOTE
Zero All Pressures
does not zero a connected ICP channel.
The ICP channel must be zeroed separately. When the
Zero ICP separately
message displays, you can zero
the ICP channel by pressing the
Zero P1
to
P8
or zero
all
module key or by selecting
Zero
from the ICP
Setup
window.
5.
Check that a zero reference has been established. Watch the pressure parameter
window for messages.
6.
Close the venting stopcock to air and open the transducer stopcock to the patient.
7.
Check that pressure numerics display on screen.
Selecting an invasive pressure channel label
One channel label can only be mapped to one channel at a time. If you select a
channel label that is already mapped to another channel, the other channel’s label
will change to the default value.
1.
Select the invasive pressure parameter window.
2.
Select the
Setup
tab.
3.
Select a channel label from the
Label
list.
Selecting the size of the invasive pressure waveform
1.
Select the invasive pressure parameter window.
2.
Select the
Setup
tab.
3.
Set the waveform scale with the
Scale
arrows.
The larger the scale value, the smaller the waveform size.
Optimizing the invasive pressure waveform scale
You can select an automatic calculation for an optimized waveform size. This size
will then be used for the local waveform, minitrend, and waveform printouts. Other
instances (e.g., information sent to the network), will use the scale selection that is as
close as possible to the upper limit of the optimized scale.
The algorithm uses the last four seconds of the waveform data to calculate the scale.
If you notice a considerable change in the waveform during that time, wait for the
waveform to stabilize and perform the operation again.
240
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...