Messages
Message
Location
Explanation
What to do
●
Identical PP modules
●
al. area
There are two or more E-PP
modules in the system.
●
Remove all but one E-PP
module.
●
Identical PT modules
●
al. area
There are two or more E-PT
modules in the system.
●
Remove all but one E-PT
module.
●
IP module in invalid
slot [1, 2, 3, 4]
●
al. area
The invasive pressure module is
in the wrong module slot.
●
Insert single parameter
pressure module in either slot
3 or 4 of the Tram-Rac 4A
housing.
●
IP's not zeroed
●
al. area
The invasive pressure lines have
not been zeroed.
●
Perform zeroing.
●
LAP sys high
/
LAP sys
low
●
LAP mean high
/
LAP mean low
●
LAP dia high
/
LAP dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
No P1 transducer
to
●
No P8 transducer
●
al. area
No transducer connected to
the channel indicated in the
message, or the sensor is faulty.
●
Connect a transducer.
●
Check the cable and
connections.
●
Replace the sensor.
●
Replace the transducer.
●
If the problem persists,
contact authorized service
personnel.
●
P1 over range
to
P8
over range
●
P1 under range
to
P8
under rang
●
al. area
The measurement value is over
or under range, or the sensor is
faulty.
●
Check the patient status.
●
Check the cables.
●
Rezero the transducer.
●
Replace the sensor.
●
Replace the transducer.
●
Replace the module.
●
If the problem persists,
contact authorized service
personnel.
●
P1 standby
to
P8
standby
●
param.
The IP channel has been set to
standby.
●
Reactivate the channel by
selecting
Activate P1
to
Activate P8
.
478
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...