Laboratory data
3.
Select
Arterial
,
Venous
, or
Other
from the
Sample Site
list.
Note that selecting
Arterial
or
Venous
affects the labels of
pH
,
PO2
,
PCO2
, and
SO2
:
●
Arterial
changes the labels to
pHa
,
PaCO2
,
PaCO2
, and
SaO2
.
●
Venous
changes the labels to
pHv
,
PvO2
,
PvCO2
, and
SvO2
.
Selecting the blood sample time for laboratory values
1.
Select
Data & Pages
.
2.
Select
Laboratory Data
.
3.
Select the
Enter Data
tab.
4.
Set the
Sample Time
with arrows.
5.
Select
Save
.
Temperature correction
In the laboratory, blood gas values are measured and calibrated at +37 °C (+99
°F). The
pH
,
PCO2
, and
PO2
values may need to be corrected to the actual patient
temperature because an increase or decrease in temperature changes the amount of
dissolved blood gas molecules and pH.
While the
Enter Data
tab of the
Laboratory Values
menu shows both the corrected
and uncorrected values, the
View
tab shows either the corrected or uncorrected
values depending on the
Temperature Correction
selection.
Formulas used to calculate the values when temperature correction is enabled are
given in the CARESCAPE Modular Monitors Supplemental Information Manual.
Selecting the type of temperature correction
1.
Select
Data & Pages
.
2.
Select
Laboratory Data
.
3.
Select the
Enter Data
tab.
4.
Select an option from the
Temperature Correction
list:
●
Laboratory
: Temperature correction has been done in the laboratory and the
values have already been corrected to patient temperature. The entered
pH
,
PCO2
, and
PO2
values are stored without adjustment and they are shown in
the
Temp corrected
column.
●
Yes
: The monitor will perform correction calculations. Select a temperature
source from the
Temperature Source
list and the monitor recalculates the
entered blood gas values corrected to patient temperature. Both the corrected
and uncorrected values are shown.
●
No
: No temperature correction is needed or performed. The entered blood
gas values are shown as such.
Entering or loading laboratory values
1.
Select
Data & Pages
.
2.
Select
Laboratory Data
.
3.
Select the
Enter Data
tab.
402
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...