Setting up the monitor before use
CAUTION
The waveform autoscaling feature automatically updates the
display from the best possible signal amplitude. Always make
sure that the waveform display scale is correctly understood
and does not lead to delayed patient treatment.
Selecting the display mode for IP waveforms
You can select the invasive pressure waveforms to be shown as individual waveforms,
or in a combined view.
1.
Select
Monitor Setup
>
Screen Setup
.
2.
B850 and B450 with the Double Video license: Select
Screen 1
or
Screen 2
tab.
3.
Select
Upper Parameter Area
.
4.
Select an option from the
Invasive Pressure Waveforms
list:
●
To view individual waveforms, select
Individual
.
●
To combine the currently displayed adjacent waveforms (2 to 4), select
Combined
. The new waveform field will use the combined height of the original
fields.
●
To combine up to four waveforms in one field, select
4invP
. The new waveform
field will use the height of two upper parameter windows.
Setting up a split screen
You can split the waveform area into two parts. The split screen option divides the
screen so that you can view graphic and/or numeric data of the chosen measurement
on the left while still having waveforms and parameter windows visible at the same
time.
1.
Select
Monitor Setup
>
Screen Setup
.
2.
B850 and B450 with the Double Video license: Select
Screen 1
or
Screen 2
tab.
3.
Select
Split Screen
.
4.
Select the type of split screen you need from the dropdown list:
●
Off
: no split screen
●
ST
shows current and reference QRS complexes and ST trends.
●
Spiro 1
is a basic view of Patient Spirometry data.
●
Spiro 2
is an enhanced view of Patient Spirometry data.
●
EEG
shows the EEG compressed spectral array (CSA).
●
AEP
shows the current auditory evoked potentials (AEP) waveforms.
●
Minitrend
shows minitrends beside waveforms.
Locked alarm and parameter settings
Some care unit and profile settings can be locked. Clinicians cannot adjust the locked
settings for the admitted patient. These settings are indicated with a lock symbol:
.
The alarm and parameter settings that can be locked vary by acquisition module.
They are set in the
Care Unit Settings
and they are password protected.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
2062971-001
CARESCAPE Modular Monitors
111
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...