CO2 with CAPNOSTAT Mainstream, CapnoFlex LF, and Dual CO2 modules
3.
Accessory port (for sample lines, nasal cannulas etc.)
4.
CapnoFlex LF module
5.
CapnoFlex LF adapter
6.
Exhaust line
Preparing the setup for CapnoFlex LF module
1.
Make sure the CapnoFlex LF adapter is attached to the Tram-Rac.
2.
Plug the CAPNOSTAT Mainstream or Dual CO2 module into the Tram-Rac.
3.
Plug the adapter cable into the CAPNOSTAT Mainstream or Dual CO2 module.
4.
Connect the sample line to the CapnoFlex LF module but not to the patient.
5.
Connect the CapnoFlex LF module into the CapnoFlex LF adapter. The message
Warming up
appears on the screen. Wait until it disappears.
6.
If the message
Calibrate sensor
appears, perform calibration.
7.
Block the open end of the sample line or nasal cannula. The message
Sample
line blocked
should appear within 60 seconds on the monitor display to ensure
that the monitor recognizes blockages.
8.
Remove the sample line from the CapnoFlex LF module and reinsert it to return
the module to normal operation.
9.
Connect the sample line to the patient’s breathing circuit.
Calibrating the CapnoFlex LF adapter
CapnoFlex LF adapter calibration should be performed whenever the message
Check/calibr. adapter
appears.
1.
Place the sample line away from all sources of CO
2
, including the patient’s exhaled
breath, your exhaled breath, and ventilator exhaust valves.
2.
With the adapter connected to the sensor, the sensor connected to the monitor,
and the pump turned on, select a gas related parameter window.
3.
Select the
Calibration
tab.
4.
Select
Start Calibrating
.
The message
Calibrated
appears on the screen after a successful calibration. If the
calibration fails, the message
Calibration error
appears instead. In this case, perform
a new calibration.
Equipment connection with Dual CO2 module
312
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...