Non-invasive blood pressure
2.
Set the
Completed NIBP Volume
.
The lower the value, the softer the tone.
Setting the NIBP display format
1.
Select the NIBP parameter window.
2.
Select the format from the
Display Format
list:
●
Sys/Dia (Mean)
: All values are shown, but the sys/dia values are shown in a
bigger font.
●
(Mean) Sys/Dia
: All values are shown, but the mean value is shown in a bigger
font.
NIBP alarms
Setting NIBP alarms
1.
Select the NIBP parameter window.
2.
Select the
Alarms
tab.
3.
Select
Systolic (SYS)
,
Mean (M)
, or
Diastolic (DIA)
pressure.
If the feature is not active, the alarm limits are greyed out. Select
Alarm On
to
set the alarms.
4.
Set the alarm limits.
Silenced NIBP alarms
The silence alarm behavior is different for NIBP than for any other parameter. Unlike
the continuously monitored parameters, NIBP is measured periodically. As a result,
silencing a physiological NIBP alarm will clear that active alarm until the next NIBP
measurement is taken. If the new measurement is outside the alarm limits, the alarm
is activated again.
NIBP recheck after alarm violation
NOTE
PDM and PSM only.
If the NIBP value exceeds the alarm limits and the NIBP alarm priority selection is
Escalating
, a new measurement takes place automatically. If the NIBP measurement
is taken manually, the recheck measurement is taken immediately after the first
measurement. When the NIBP measurement is taken automatically, the recheck
measurement is delayed by 30 seconds before the second measurement is taken.
NIBP measurement description
The acquisition modules use oscillometric technology to acquire NIBP. Oscillometry
is the most commonly used means of indirect blood pressure measurement in
automated devices. It is based on the principle that pulsatile blood flow through an
artery creates oscillations of the arterial wall.
Oscillometric devices use a blood pressure cuff to sense these oscillations that appear
as tiny pulsations in cuff pressure. By measuring and analyzing at various cuff
228
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...