Cardiac output
1.
Select the cardiac output parameter window.
2.
Select the
Setup
tab.
3.
Select the radio button for
Automatic
measurement type.
4.
Verify that the catheter settings are correct.
5.
Select the
Measurement
tab.
6.
Complete the following:
a. PDM and TRAM: Get ready to inject the injectate solution when the message
Inject When Ready
appears.
b. E-modules: When the message
Press Start C.O. Serial
appears, select
Start
C.O. Serial
.
7.
Inject the injectate solution smoothly within 4 to 5 seconds.
8.
The message
Measuring
displays, followed by the message
Please wait
until
the calculation is completed.
9.
Observe the washout curve displayed on the screen. The message
C.O. Complete
displays after the C.O. determination has been made, followed shortly by the
message
Please wait
.
The curve disappears from the screen when the next measurement cycle can
start.
10. To take another C.O. measurement, wait for this message to display before
injecting the injectate:
●
PDM and TRAM:
Inject When Ready
●
E-modules:
Inject now!
Taking a manual C.O. measurement
Measuring C.O. using the manual mode allows you to determine when to begin the
injection procedure. This mode may be preferred for patients with extreme blood
temperature fluctuations, or when the automatic mode is unable to establish a stable
baseline.
Holding the injectate syringe by the plunger, not the barrel, improves measurement
accuracy.
1.
Select the cardiac output parameter window.
2.
Select the
Setup
tab.
3.
Select the radio button for
Manual
measurement type.
4.
Select the
Measurement
tab and verify that the catheter settings are correct.
5.
Select
Start C.O.
With E-modules, you can also use the
Start C.O.
module key.
6.
When the
Inject now!
message appears, inject the injectate solution smoothly
within 4 to 5 seconds.
7.
The message
Measuring
displays, followed by the message
Please wait
until
the calculation is completed.
2062971-001
CARESCAPE Modular Monitors
265
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...