Alarms
Technical alarms’ deactivation with the pause
audio key
Certain technical alarms can be deactivated with the pause audio key. The
Alarm
Deactivation
setting is configured in the
Care Unit Settings
and it is password
protected.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Technical alarms for which the deactivation with the pause audio key can be allowed
are:
●
ECG Leads Off
●
Art disconnect
●
ABP disconnect
●
Fem disconnect
●
UAC disconnect
●
SpO2 Probe Off
Apnea alarms’ deactivation with the pause audio
key
Apnea alarms can be deactivated with the pause audio key if the
Allow alarm
deactivation with the Audio Pause key for:
setting
Apnea (CO2/Imped)
is enabled in
the
Care Unit Settings
. This setting is password protected.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Pause audio with combination monitoring
When using combination monitoring, the pause audio behavior is the following:
●
If the telemetry transmitter is in pause audio state, also the monitor will be in the
same state. You can cancel the pause audio at the bedside monitor by selecting
the pause audio key. This will not affect the telemetry device.
●
If the monitor’s own pause audio state ends before the telemetry transmitter’s
audio pause, the monitor will re-enter pause audio.
●
The pause audio started by the telemetry transmitter will end also at the monitor
when the transmitter’s pause audio ends.
Breakthrough alarms
The breakthrough alarms feature allows pre-defined and user-selectable alarms to
“break through” (interrupt) an
All Alarms Audio Off
or a 2 or 5 minute audible alarm
pause condition.
The
FiO2 low
,
EtO2 low
,
FiN2O high
, and
Ppeak high
alarms will always break through
when escalated to or activated at high priority alarm condition regardless of the
All
Alarms Audio Off
selection or any alarm pausing.
2062971-001
CARESCAPE Modular Monitors
139
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...