Invasive pressures
NOTE
PDM and TRAM may use either the automated wedge
program or a manual measurement to determine the PCWP
value. The PA wedge algorithm requires a 30% change in
waveform size between the PA and Wedge waveform in
order to initiate the automated program. If the algorithm fails
to distinguish between the waveforms, you should use the
manual measurement mode.
Showing the PCWP value in the PA window
1.
Select the PA invasive pressure parameter window.
2.
Select the
Setup
tab.
3.
Select
Show PCWP
.
Taking a manual PA wedge measurement
NOTE
PDM and TRAM only.
1.
Select the PA invasive pressure parameter window.
2.
Select the
Setup
tab.
3.
Select
Wedge
.
4.
Select
Mode: Manual
.
5.
To record a realtime PA wedge waveform during analysis, select
Print PA
Waveform
. To stop printing, select
Stop Printing
or
Cancel Printing
.
6.
Inflate the catheter balloon when the
Manually "Freeze / Adjust" when ready
message displays.
7.
Select
Freeze/Adjust
once the PCWP waveform is displayed. The
Wedge
Complete
message displays.
8.
Deflate the balloon.
9.
To adjust the PA wedge value, move the cursor up or down with the
PCWP
/ Cursor
arrows.
10. To save the PCWP value, select
Confirm Wedge
.
The saved PA wedge value displays in the parameter window and is stored in
trends and hemodynamic calculations.
11. To print a PCWP report, select
Print PA Report
. To stop printing, select
Stop
Printing
or
Cancel Printing
.
The PA wedge report contains 20 seconds of waveform data displayed at a
waveform speed of 12.5 mm/s.
Taking an automated PA wedge measurement
1.
Select the PA invasive pressure parameter window.
2.
Select the
Setup
tab.
3.
Select
Wedge
.
4.
Select
Mode: Auto
.
2062971-001
CARESCAPE Modular Monitors
247
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...