Airway gases
Using the CARESCAPE respiratory modules and
compact airway modules for O₂ measurement
Selecting the O
2
scale
If the difference between FiO
2
and EtO
2
is more than 6%, change the O
2
scale.
1.
Select a gas related parameter window.
2.
Select the
O2
tab >
Setup
.
3.
Select an option from the
Scale
list.
Selecting the O
2
sweep speed
This selection affects the waveform.
1.
Select a gas related parameter window.
2.
Select the
O2
tab >
Setup
.
3.
Select an option from the
O2 Sweep Speed
list. The options are
0.625 mm/s
,
6.25 mm/s
,
12.5 mm/s
, and
50 mm/s
.
The smaller the value, the slower the sweep speed.
Setting O
2
alarms
1.
Select a gas related parameter window.
2.
Select the
O2
tab >
Alarms
.
3.
Check that the required alarm (
EtO2
or
FiO2
) is on, and set its high and/or low
limit values.
Using the CARESCAPE respiratory modules
and compact airway modules for AA and N₂O
measurement
Selecting the agent scale
Every anesthetic agent has its own default scale that the monitor uses when detecting
the agent. You can change the scale of an agent if the amount used is higher than
the default scale. Default scales are given in the CARESCAPE Modular Monitors
Supplemental Information Manual.
1.
Select a gas related parameter window.
2.
Select the
Agent/N2O
tab >
Setup
.
3.
Select an option from the
Agent Scale
list.
Selecting the agent sweep speed
This selection affects the waveform.
294
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...