Messages
Message
Location
Explanation
What to do
●
SV Tachy
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
●
Telemetry patient: No
12SL available
●
param.
12SL is not available with
telemetry.
●
If 12SL is required, use
a module that is directly
connected to the monitor
instead of a telemetry
transmitter.
●
Tachy
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
●
Telemetry battery
empty
●
al. area
The telemetry transmitter’s
battery has no charge left.
●
Change the battery.
●
Trigeminy
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
●
TTX Off Network
●
al. area
No parameter or waveform
data has been received from the
telemetry transmitter for more
than 5 seconds.
●
Wait to see if the condition
resolves. If the problem
persists, contact authorized
service personnel.
●
V Brady
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
●
V Fib/V Tach
●
al. area,
param.,
wavef.
Physiological alarm.
●
Check the patient status.
●
V lead A is not V1 or V
lead B is not V5
●
report
Va lead is not V1 and/or Vb lead
is not V5.
●
Check the settings and
correct if necessary.
●
V leads off
●
report
All V leads are off.
●
Connect the V leads to the
patient.
●
V Tach
●
al. area,
param.,
wavef.
Physiological alarm.
●
Check the patient status.
●
V2/C2 lead off
●
V3/C3 lead off
●
V4/C4 lead off
●
V5/C5 lead off
●
V6/C6 lead off
●
V/C lead off
●
Va/Ca lead off
●
Vb/Cb lead off
●
wavef.
An electrode is disconnected.
●
Check the electrodes.
●
VT>2
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
2062971-001
CARESCAPE Modular Monitors
467
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...