Trends
4.
Select a value from the
Minitrend Length
list.
●
The 1 minute and 2 minute minitrends (high-resolution license) are updated
every 2 seconds.
●
The 5 minute and 10 minute minitrends are updated every 10 seconds.
●
Other lengths are updated once a minute.
Selecting high-resolution contents to minitrend
If the high-resolution license is enabled and the
10 min wide
is 1 minute normal or
2 minutes wide, you can select CO
2
and impedance respiration (10Hz)
Compressed
Waveform
minitrends or
Resp Rate
minitrends.
1.
Select
Monitor Setup
>
Screen Setup
.
2.
B850 and B450 with Double Video license: Select
Screen 1
or
Screen 2
tab.
3.
Select the
Split Screen
tab.
4.
Select
Minitrend
from the
Show
list.
a. Select
Resp Rate
from the
CO2 Minitrend
or from the
Imp. Resp Minitrend
list.
b. Select
Compressed Waveform
from the
CO2 Minitrend
or from the
Imp. Resp
Minitrend
list.
Removing minitrend from the screen
1.
Select
Monitor Setup
>
Screen Setup
.
2.
B850 and B450 with the Double Video license: Select
Screen 1
or
Screen 2
tab.
3.
Select the
Split Screen
tab.
4.
Select the
None
from the
Show
list.
Time change during a patient case
CAUTION
System time changes will result in time differences between
stored and realtime data.
Time adjustment is allowed during a patient case if the monitor is configured to the
CARESCAPE Network. When the time is adjusted, the monitor shifts the timestamps of
continuous trend data and of discrete data, except NIBP and wedge measurements.
After time adjustment continuous and discrete data cannot be compared to each
other because their timestamps no longer match.
2062971-001
CARESCAPE Modular Monitors
423
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...