Messages
Message
Location
Explanation
What to do
●
In vivo poor signal
●
param.
The measurement cannot be
performed because the signal is
too weak.
●
Check connections and
catheter.
●
Change optical module and
recalibrate.
●
If the problem persists, the
catheter could be faulty;
replace the catheter.
●
Insufficient signal
●
param.
The measurement cannot be
performed because the signal is
too weak.
●
Check the patient status.
●
Check connections and
catheter.
●
If the problem persists,
contact authorized service
personnel.
●
Intensity shift
●
param.
The measurement cannot be
performed because the signal
intensity changed.
●
Check the patient status.
●
Check connections and
catheter.
●
If the problem persists,
contact authorized service
personnel.
●
Low Intensity
●
param.
Catheter floated or up against
vessel wall.
●
Check the optical module and
connections.
●
Check the catheter
placement.
●
Only authorized medical
personnel should adjust
catheter placement.
●
Low Light
●
param.
Catheter floated or up against
vessel wall.
●
Check the optical module and
connections.
●
Check the catheter
placement.
●
Repeat in vivo calibration.
●
No Light
●
param.
Interfaced device should be
checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
Re-calibrate SvO2
●
al. area,
param.
The calibration is over 24 hours
old.
●
Perform in vivo calibration.
●
Start SvO2
●
param.
Displays when the catheter has
been inserted into the patient
after in-vitro calibration.
●
Select the
Start SvO2
option
from the SvO
2
menu.
●
SvO2 cable off
●
al. area
The cable is disconnected from
the module.
●
Re-connect the cable to the
module.
●
SvO2 faulty cable
●
al. area,
param.
The optical module has failed.
●
Check the optical module and
connections.
●
Replace the optical module.
2062971-001
CARESCAPE Modular Monitors
487
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...