Pulse oximetry
1.
Select the SpO
2
parameter window.
2.
Select the
SpO2
tab.
3.
Adjust the volume with the
Beat Volume
arrows.
Variable beat tone
You can configure a variable beat tone through
Monitor Setup
>
Default Setup
>
Care
Unit Settings
>
Parameters
>
Variable Beat Tone
. This setting is password protected.
If it set to
All beat sources
, the SpO
2
saturation affects all beep sounds including ECG
and IP when the SpO
2
measurement is available: beep frequency changes according
to increasing and decreasing SpO
2
values. If the setting is set to
Only SpO2
, other beep
sounds are not affected by the changing SpO
2
values.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Masimo SET data averaging and updating
For Masimo SET technology, when using the default averaging time of 8 seconds,
there is a maximum data-averaging signal processing time of 10 seconds from real
time plus an additional delay of 2 seconds to update the displayed waveform. Audible
alarms are delayed until an alarm limit violation occurs for at least 5 seconds.
Selecting the SpO
2
averaging time
NOTE
PSM, E-MASIMO, and PDM and TRAM with Masimo technology
and Masimo sensors only. The primary SpO
2
measurement
only.
You can have an average of the SpO
2
measurement on screen instead of the beat to
beat values, and you can select how many seconds are used for this averaging:
2 s
,
4 s
,
8 s
,
10 s
,
12 s
,
14 s
, or
16 s
.
1.
Select the SpO
2
parameter window.
2.
Select the
SpO2
tab.
3.
Select the number of seconds from the
Averaging
list.
Selecting the Masimo SpO
2
sensor sensitivity level
NOTE
SpO
2
modules with Masimo technology and Masimo sensors
only.
1.
Select the SpO
2
parameter window.
2.
Select the
SpO2
tab.
3.
Select the appropriate
Sensitivity
radio button:
●
Use the
Normal
sensitivity setting for normal patient monitoring purposes.
●
Use the
Maximum
sensitivity setting for improved poor perfusion performance
and for faster tracking of rapid SpO
2
saturation changes.
Using the
Maximum
sensitivity setting delays the
SpO2 probe off
detection alarm.
210
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...