Mixed venous oxygen saturation (SvO₂)
Using the SvO₂ measurement
SvO
2
calibration in vitro
1.
Module E-COPSv
2.
Optical module
3.
Optical connector
4.
Swan-Ganz thermodilution catheter
Calibrating a new SvO
2
catheter in vitro
NOTE
Do not perform in vitro calibration if the catheter has been
flushed. Using a wet catheter and calibration cup results in an
inaccurate calibration.
Always perform in vitro calibration with a new catheter before removing it from its
package. Follow the
Advice to User
information displayed in the
Calibration
window
to guide you through the calibration steps.
1.
Connect the optical module to the module and let it warm up for 20 minutes.
2.
Aseptically expose the catheter's optical connector.
3.
Connect the catheter to the optical module.
4.
Select the SvO
2
parameter window.
5.
Select the
Calibration
tab.
6.
If replacing an existing catheter, select
New Catheter
.
7.
Select
In Vitro Calibration
>
Calibrate
to perform in vitro calibration.
8.
Select
Start SvO2
to complete the in vitro calibration.
9.
Insert the catheter into the patient.
Recalling a previous in vitro calibration
After completing a successful initial calibration, you can recall the previous in vitro
calibration measurement when:
●
The optical module is connected and has not been calibrated since being
connected.
2062971-001
CARESCAPE Modular Monitors
277
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...