Messages
Message
Location
Explanation
What to do
●
PEEP low
●
PEEPe high
/
PEEPe
low
●
PEEPi high
/
PEEPi low
●
PEEPtot high
/
PEEPtot low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Ppeak high
/
Ppeak
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Saving Loop
●
al. area
A loop is being saved.
●
Wait until the message
disappears.
●
Scale changed
●
wavef.
The
Auto
scaling mode has
changed the
Flow
,
Paw
or
Vol
scale.
●
Wait until the message
disappears.
●
Service gas module
●
Service gas module
- and specific error
indication
●
al.area,
param.,
wavef.
The measuring sensor is
inoperative or the temperature
in the module has increased.
●
Contact authorized service
personnel.
●
TVexp low
●
al. area
Measured value is outside or
equal to the alarm limit.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Vent: Apnea
●
al. area
No breathing detected.
●
Check the patient status.
●
Check the interfaced device.
●
Vent: Check
ventilator
●
al. area
The interfaced device needs to
be checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
Vent: Connecting
●
al. area
Interfaced device is being
connected to the monitor
through Unity Network Interface
Device (ID).
●
No action required.
●
Vent: Disconnect
●
al. area
Interfaced device needs to be
checked.
●
Check the patient status.
●
Check the ventilator tubing
and patient connection.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
494
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...