Messages
Message
Location
Explanation
What to do
●
HR high
/
HR low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Irregular
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
●
LA/L lead off
●
LL/F lead off
●
RA/R lead off
●
RL/N lead off
●
wavef.
An electrode is disconnected.
●
Check the electrodes.
●
Lead changed
●
wavef.
The monitor automatically
switches the ECG1 waveform
selection to a measurable ECG
Lead (I, II, III, aVR, aVL, aVF,
V1, V2, V3, V4, V5 or V6) if the
current ECG1 waveform is not
measurable.
●
Note that the ECG waveform
changes according to the
lead it is measured from.
Check the lead.
●
Leads off
●
al. area,
wavef.
One or more of the connected
electrodes is disconnected and
arrhythmia detection is not
possible.
●
ECG: Check the connections.
●
Learning
●
wavef.
ST algorithm is in learning phase,
message shown e.g. when ECG
measurement is started.
●
No action required.
●
Missing beat
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
●
Multifocal PVCs
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
●
No 10 or 6 lead cable
●
report
No 6- or 10-lead ECG cable is
attached.
●
Connect a 6- or 10-lead cable.
●
No 12RL license
●
report
No 12RL 12 lead ECG license.
●
Contact your GE
representative to purchase
this feature.
●
No 12SL license
●
report
No 12SL ECG with ACI-TIPI
license.
●
Contact your GE
representative to purchase
this feature.
●
No telemetry
●
al. area
The server cannot communicate
with the telemetry transmitter.
●
Make sure the transmitter is
not out of range.
●
Check the telemetry
transmitter’s battery.
●
Wait to see if the problem
resolves. If the problem
persists, contact authorized
service personnel.
2062971-001
CARESCAPE Modular Monitors
465
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...