Neuromuscular transmission
ElectroSensor and electromyography (EMG)
Measurement with ElectroSensor is based on electromyography. Three electrodes are
placed on the muscle, and the innervating nerve is stimulated. The recorded electrical
activity of the muscle is inversely proportional to the degree of block.
Stimulation modes
●
Train of four, TOF: Recommended for most cases. It is also the default setting.
●
Double burst stimulation, DBS: Useful when using the MechanoSensor. It enables
better visual observation of the fading in responses.
●
Post tetanic count, PTC: Used for estimating the relaxation level with tetanic
stimulation.
●
Single twitch, ST: Single twitch mode is practical when using depolarizing relaxants:
in these cases, TOF% does not give any additional information about the patient
status.
How to interpret the NMT values
When neuromuscular block deepens, different stimulation modes may be needed to
assess the relaxation status. The following table describes the depth of relaxation.
100
TOF%
20
4
Count
0
10
PTC
0
Light
RELAXATION METER
Deep
Train of four (TOF) mode
In TOF stimulus mode, four stimulation pulses are generated at 0.5 second intervals.
The response is measured after each stimulus and the ratio of the fourth to the first
response of the TOF sequence is calculated, resulting in TOF%.
With the ElectroSensor, T1% is displayed. If the reference is successfully found, a scale
is also included. Scale markers represent 0%, 25%, 50%, 75% and 100% reference
values. When no reference is available, no T1% value is displayed and the bars are
not scaled.
When relaxation deepens, the TOF% declines until the fourth response disappears and
no TOF% is available. The degree of neuromuscular block is then estimated from the
number of responses, the count, which represents the number of responses detected
to the four stimuli. The fewer the responses, the deeper the relaxation.
Number of responses, count
Neuromuscular block
Muscle power (% of control)
1
95%
5%
2
90%
10%
3
85%
15%
4
75%
25%
Double burst stimulation (DBS) mode
DBS consists of two separate bursts. Each burst consists of three consecutive pulses
at a 50 Hz frequency. The response ratio of the second to the first burst is calculated,
resulting in DBS% (equivalent to TOF%).
2062971-001
CARESCAPE Modular Monitors
369
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...