Alarms
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
When audible alarms are turned off:
●
All audible alarms are turned off except for any high priority alarms configured
to break through the audio off setting.
●
The audio off bell icon displays in the upper left corner of the display screen.
●
The alarm audio pause/off area of the alarm light is solid blue when audible alarms
are paused or when audio off is selected for an alarm group.
Turning audible alarms on/off
You can turn on/off the audible physiological alarm tones for an alarm group or for
all alarms.
1.
Select
Alarm Setup
from the monitor’s main menu.
2.
Select the
Audible & Visual
tab.
3.
Select an alarm group. Choices are:
●
None
: No audible alarms are turned off.
●
Apnea Audio Off
: Turns off audible alarms for apnea, EtCO
2
, FiCO
2
, respiration
rate, Ppeak low, PEEPe, PEEPtot, PEEPi, and MVexp limit alarms.
●
ECG Audio Off
: Turns off audible alarms for all HR and PR source limit and
arrhythmia alarms.
●
Apnea & ECG Audio Off
: Turns off audio alarms for all HR and PR source limit,
arrhythmia, apnea, EtCO
2
, FiCO
2
, respiration rate, Ppeak low, PEEPe, PEEPtot,
PEEPi, and MVexp limit alarms.
●
All Alarms Audio Off
: Turns off all audible alarms except some high priority
alarms defined as breakthrough alarms.
4.
To turn on all audible alarms again, select
Activate All Audible Alarms
, or select
None
as instructed above.
NOTE
If alarms are turned off for any of the defined alarm groups
and an alarm occurs within the alarm group, a beep tone will
sound every 2 minutes as a reminder that alarms are turned
off.
NOTE
France only: A reminder beep tone sounds every 2 minutes
when the
Audible & Visual
>
None
setting is not selected.
Pause audio behaviors
Selecting the pause audio key results in different alarm behaviors depending on
whether the alarms are active and/or latched or not. Acknowledging or pausing audio
alarms does not affect other alarm indicators. They will still continue indicating alarms.
When the monitor is on the network, alarms can also be paused and acknowledged at
the central station.
2062971-001
CARESCAPE Modular Monitors
137
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...