Messages
Message
Location
Explanation
What to do
●
Wrong cable
●
param.
You have connected a
MasimoSet cable to the PDM
Nellcor, or a Nellcor OxiMax
cable to the PDM Masimo.
●
Check and use the correct
cable.
●
Wrong cable. Use
Masimo Set
●
al. area,
param.
You have connected a Nellcor
OxiMax cable to the PDM
Masimo.
●
Use a MasimoSet cable.
●
Wrong cable. Use
Nellcor OxiMax
●
al. area,
param.
You have connected a
MasimoSet cable to the PDM
Nellcor.
●
Use a Nellcor OxiMax cable.
Messages related to NIBP measurement
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
●
report = report view
●
wavef. = waveform area
Message
Location
Explanation
What to do
●
Air leakage
●
param.
Loose cuff or cuff hose.
●
Check the cuff and cuff hose.
●
Artifacts
●
param.
Measurement is disturbed by
artifact (e.g. patient movement,
shivering, deep breathing,
marked arrhythmia or irregular
beats).
●
Calm the patient and retry.
●
Calibrated
●
param.
Channel calibrated successfully.
●
Wait until the message
disappears before starting a
measurement.
●
Calibrating
●
param.
Calibration of a channel is in
progress.
●
No action required.
●
Calibration switch ON
●
param.
The calibration switch is on.
●
Contact authorized service
personnel.
●
Call service
●
param.
Technical fault.
●
Contact authorized service
personnel.
472
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...