CO2 with CAPNOSTAT Mainstream, CapnoFlex LF, and Dual CO2 modules
4.
Place the sensor on the cell marked
0
and select
Start Zeroing
.
5.
After the message
Zeroed
appears, remove the sensor from the
0
cell and attach
it to the cell marked
REF
.
The measured EtCO
2
value appears in the
Connect sensor to REF cell
window. If
the value stays between 36 and 40 mmHg for three seconds during the 20-second
calibration, the calibration is successful and the message
Calibrated
appears. If the
calibration fails, the message
Recalibrate
appears instead. In this case, perform
a new calibration.
Calibrating the CAPNOSTAT Mainstream adapter
Adapter calibration should be performed whenever the message
Check/calibr.
adapter
appears.
1.
Place the sensor and adapter away from all sources of CO
2
, including the patient’s
exhaled breath, your exhaled breath, and ventilator exhaust valves.
2.
With the adapter connected to the sensor and the sensor connected to the
monitor (and, when used, all sidestream tubing attached and the pump turned
on), select the gases parameter window.
3.
Select the
Calibration
tab.
4.
Select
Start Calibrating
.
The message
Calibrated
appears on the screen after a successful calibration. If the
calibration fails, the message
Calibration error
appears instead. In this case, perform
a new calibration.
Equipment connection with CapnoFlex LF module
For intubated and non-intubated patients.
1.
CAPNOSTAT Mainstream or Dual CO2 module
2.
Adapter cable
2062971-001
CARESCAPE Modular Monitors
311
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...