Patient Spirometry
2.
Select
Loops
.
3.
Select
Print Loop
.
NOTE
With compact airway modules, you can also print the loop by
pressing the
Print Loop
module key.
Setting Paw alarm limits
You can set the limit alarms on or off, and adjust their activation limits according
to your needs.
1.
Select the Patient Spirometry parameter window.
2.
Select
Paw Alarms
.
3.
Set the alarm limits: PEEPtot (OR and PACU software packages), PEEPi and PEEPe
(other software packages), Ppeak and MVexp.
NOTE
If the feature is not active, the alarm limits are greyed out.
Select
Alarm On
to set the alarms.
Setting MV/Vent alarm limits
You can set the limit alarms on or off, and adjust their activation limits according
to your needs.
1.
Select the Patient Spirometry parameter window.
2.
Select
MV/Vent Alarms
.
3.
Set the alarm limits.
NOTE
If the feature is not active, the alarm limits are greyed out.
Select
Alarm On
to set the alarms.
NOTE
MV/Vent Alarms
cannot be selected off except when an
interfaced device is used.
NOTE
To enable apnea and technical alarms from the ventilator,
select the
Ventilator Apnea and Technical Alarms
check
box.
Patient Spirometry measurement basics
Patient Spirometry measurement description
The CARESCAPE respiratory modules E-sCOV and E-sCAiOV, and the compact airway
modules E-COV, E-COVX, E-CAiOV, and E-CAiOVX measure airway pressures, flow,
volumes, compliance and airway resistance breath by breath at the patient's airway.
All parameters are measured through D-lite(+)/Pedi-lite(+) flow sensor placed at the
patient's airway.
328
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...