ECG
■
Pace 2
: Minimizes the possibility of counting pacemaker artifact as QRS
complexes during asystole.
■
Pace 1
: Does not minimize the possibility of counting artifact as a QRS complex
during asystole. If the monitor resets or is discharged, or the profile is changed,
and the monitor is set to
Pace 1
, the monitor automatically changes to the
Pace 2
setting.
NOTE
TRAM version 12 or later and telemetry with ApexPro
4.1 or later with enhanced pacemaker detection will
automatically detect pacemakers with either
Pace 1
or
Pace 2
. Selecting one or the other makes no difference in
this respect.
Pacemaker detection troubleshooting
Problem
Solution
How does activating pacemaker detection impact
monitoring?
●
Beats that would otherwise be classified as
ventricular are instead classified as V-paced if a
ventricular pacemaker event is detected.
●
Residual pacemaker energy that might otherwise
appear in the ECG is removed, and a pacemaker
enhanced spike is placed in the ECG.
●
On the ECG waveform, pacemaker detection
is indicated by uniform, upright pacemaker
enhancement spikes in the ECG data, both
displayed and graphed.
How can pacemaker detection be improved?
●
Possible problems include:
■
Heart rate double counting.
■
Inaccurate alarms for low heart rate or asystole.
■
Pacemaker spikes not recognized by the
software.
■
False PVC detections and arrhythmia alarms.
●
Possible solutions include:
■
Relearn arrhythmia.
■
Re-prepare the patient skin, replace the
electrodes, and adjust the electrode placement.
■
Try an alternate electrode placement.
■
Try single-lead analysis, if available.
■
E-modules, TRAM, combination monitoring:
Switch to another pacemaker detection mode.
Why is the monitor double-counting the heart rate,
alarming for a low heart rate, or not detecting
pacemaker spikes?
The monitor is not detecting pacemaker activity.
Causes may include:
●
PDM, TRAM, combination monitoring: The
pacemaker detection program is turned off. Turn it
on, reprep the skin and reposition the electrodes if
necessary. Relearn ECG.
●
The pacemaker signal is too weak for the monitor
to detect.
174
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...