Messages
Message
Location
Explanation
What to do
●
Replace sensor
●
param.
The sensor date has expired, the
sensor has been used too many
times, or the validity time for the
sensor cannot be determined.
●
Replace the sensor.
●
Sensor check failed
●
param.,
wavef.
The sensor has not passed the
impedance check.
●
Check sensor placement and
attachment.
●
Press each electrode in the
sensor.
●
Replace the sensor.
●
Testing DSC
●
param.,
wavef.
The BISx test has been activated.
●
Wait until the check is
completed.
Messages related to TC measurement
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
●
report = report view
●
wavef. = waveform area
Message
Location
Explanation
What to do
●
Calibrating
●
param.
Calibration is in progress.
●
No action required.
●
Calibration error
●
param.
Unsuccessful calibration.
●
Authorized service personnel
should repeat the calibration
procedure.
●
Connecting
●
param.
Connection to the interfaced
device is being established.
●
No action required.
●
Ext device error
●
param.
Interfaced device should be
checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
No ext device
●
param.
Interfaced device should be
checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
Not calibrated
●
param.
Measurement needs to be
calibrated.
●
Perform calibration.
●
Ready
●
param.
Measurement is ready.
●
No action required.
2062971-001
CARESCAPE Modular Monitors
503
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...