Starting and ending monitoring
●
Erase PDM Data
/
Erase Tram Data
: This selection erases the patient information
and data from the module.
How to continue monitoring when a case is active/patient
is admitted
When you connect a PDM or TRAM that contains an MRN that matches the MRN
entered on the monitor, the patient information and data is automatically loaded
from the PDM or TRAM to the monitor and a case is automatically started/patient is
admitted.
When you connect a PDM or TRAM that contains patient information but the MRN
does not match the MRN entered on the monitor, the
Continue
menu appears.
NOTE
When data is combined either automatically or manually,
only the data after the last module disconnection is loaded.
To load all trends, discharge the patient from the monitor
and make sure that the PDM/TRAM is not connected to the
monitor. Then reconnect the module.
The
Continue
menu has two informative fields on the top: the
Patient in the monitor
and the
Patient in the PDM
or
Patient in the Tram
. These fields show the MRN and
Name of the Patient information if there is an active case/admitted patient. Select one
of the following to continue monitoring:
●
Load PDM Data
/
Load Tram Data
: This selection erases the patient data from the
monitor and loads the data from the module.
●
Continue Current
: Monitoring will continue with the patient data from the monitor.
If the patient identification is not the same on the acquisition module and the
moniotor, this selection erases the patient data from the monitor and continues
with the patient currently on the module.
●
Erase PDM Data
/
Erase Tram Data
: This selection erases the patient data from
the module.
●
Combine Data
: This selection combines the patient data from the module with
that of the monitor even if the patient identification is different or it has not been
entered. Use this selection if no MRN has been entered or you know for certain that
you will continue monitoring the same patient. This selection may also be useful
if there has been a typing mistake or some other minor error when entering the
patient identification. Any information available in one device and not in the other
will overwrite the missing information. Be careful when using this selection. If you
are not absolutely certain that it is the same patient on the monitor and module, do
not combine the data.
●
Discharge
: This selection deletes the patient data from the module and from the
monitor.
If you connect a PDM or TRAM module and there is no patient information or data in
the module or in the monitor, the monitoring does not start and the
Continue
menu
does not appear.
When you select to erase data or discharge a patient in the
Continue
menu, the
patient’s vital signs cannot be observed during the erase/discharge cycle time.
If the
Continue
menu is open when a request for time adjustment is received, the
adjustment will be delayed until the menu is closed and data has been loaded.
124
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...