Airway gases
●
Make sure that the gas sampling line is properly connected to the water trap and
the water trap is properly connected to the airway gas module. Gas leaks in these
connections may dilute the gas sample from the patient circuit, thus resulting in
erroneous gas readings. During normal operation, all sampled gas flows out of the
sample gas outlet. Room air is used as reference gas for the oxygen measurement
and it is mixed with the sampled gas. The sampled gas is diluted by room air so
that the fraction of room air in the exhaust gas is about 20%.
●
Compact airway modules: The message
Sample line blocked
may result if you
attach the sampling line to the water trap after the monitor has completed the
self-check for the module. Attach the sampling lines to the water trap before
turning on the monitor.
Airway gases measurement setup
Airway gases equipment to patient connections with
CARESCAPE respiratory modules
1.
CARESCAPE respiratory module
2.
Gas sample, gas sampling line connector on the water trap
3.
Gas sampling line
4.
Gas sampling line connector on the airway adapter; place the connector upwards
5.
Airway adapter with sampling line connector
6.
Heat and moisture exchanger with filter (HMEF) (optional)
284
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...