Messages
Message
Location
Explanation
What to do
●
Inject now!
●
C.O.
menu
A prompt text during the
Manual
C.O. measurement.
●
Inject the injectate solution
smoothly within 4 to 5
seconds.
●
Inject When Ready
●
C.O.
menu
A prompt text during the
Automatic
C.O. measurement.
●
Inject the injectate solution
smoothly within 4 to 5
seconds.
●
Irregular curve
●
C.O.
menu
Thermodilution waveform is
irregular.
●
Perform a new measurement.
●
Irregular HR
●
C.O.
menu
HR measurement values are
irregular.
●
Check the patient status.
●
Measuring
●
C.O.
menu
Determining the C.O. value.
●
Wait until the message
disappears.
●
No catheter
●
C.O.
menu
No catheter connected.
●
Connect a catheter.
●
No Comp Constant.
Check C.O. setup
●
C.O.
menu
The computation constant is
not available for the selected
catheter, injectate volume, and
injectate probe combination.
●
Check the C.O. setup
selections.
●
No ext device
●
param.
Interfaced device should be
checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
No injectate temp
probe
●
C.O.
menu
No injectate temperature probe
connected.
●
Connect an injectate
temperature probe.
●
No HR for REF
●
C.O.
menu
REF measurement cannot be
done.
●
Check that the patient’s heart
rate is being monitored.
●
Add HR measurement if
needed.
●
No module
●
C.O.
menu
No C.O. module connected.
●
Connect a C.O. module to
measure cardiac output.
●
Noisy baseline
●
C.O.
menu
Changes in patient’s blood
temperature affect C.O.
measuring.
●
Check the patient status.
●
Check the blood temperature
connector.
●
Please wait
●
C.O.
menu
A failed, canceled, or stopped
C.O. determination and the
module not yet ready for the
next measurement.
●
Wait until the message
disappears.
●
Press Start C.O.
●
C.O.
menu
Prompt text during the
Manual
measurement mode.
●
Proceed with the
measurement by selecting
Start C.O.
With E-modules,
you can also use the
Start
C.O.
module key.
●
Press Start C.O. Serial
●
C.O.
menu
Prompt text during the
Automatic
measurement
mode.
●
Proceed with the
measurement by selecting
Start C.O. Serial
.
484
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...