Messages
Message
Location
Explanation
What to do
●
Lab data interface
failed
●
al. area
Unity Network Interface Device
(ID): There is a communication
failure or error between the
Unity Network Interface Device
(ID) and the interfaced device.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
Mark xxx
●
where xxx = snapshot
sequence number
●
alarm
area
A snapshot has been taken
manually.
●
No action required.
●
Snapshot created
●
al. area
A snapshot has been created.
●
No action required.
●
Snapshot memory
full. Oldest snapshot
erased.
●
al. area
You are trying to save a snapshot
but the memory capacity is full.
●
No action required.
●
ST snapshot created
●
al. area
ST snapshot created.
●
No action required.
●
ST snapshot memory
full. Oldest ST
snapshot erased.
●
al. area
You are trying to save an ST
snapshot but the memory
capacity is full.
●
No action required.
●
Sweep Speed
Changed
●
wavef.
Sweep speed of the realtime
waveform has changed.
●
No action required.
●
Waveform available
only for 2min and
4min time scales
●
wavef.
High Resolution trends license is
configured for HR, Resp, SpO
2
,
CO
2
and MAP.
●
Change trend resolution to 2
minute or 4 minute time scale
for the affected parameter.
Messages related to various situations
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
●
report = report view
●
wavef. = waveform area
Message
Location
Explanation
What to do
●
ADT server
communication
failure
●
al. area
An error occurred when trying
to search for patient on the ADT
server.
●
Check the network
connectivity.
●
Retry loading from the
network.
●
If the problem persists,
contact authorized service
personnel.
●
Alarm setup changed
remotely
●
al. area
The alarm setup is retrieved from
the telemetry server or central
station.
●
Check the alarm settings and
adjust if necessary.
2062971-001
CARESCAPE Modular Monitors
505
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...