CO2 with CAPNOSTAT Mainstream, CapnoFlex LF, and Dual CO2 modules
1.
Airway adapter to patient endotracheal tube
2.
Mainstream adapter
3.
Mainstream sensor (upright position)
4.
Reference cell
5.
Zero cell
6.
Sensor cable
7.
Module
Preparing the setup for CAPNOSTAT Mainstream module
For intubated patients.
1.
Snap the airway adapter on to the mainstream sensor.
2.
Connect the sensor to the module. The message
Warming up
is displayed on the
monitor screen. Wait until it disappears.
3.
Confirm that the monitor’s atmospheric pressure setting is correct.
4.
Check the monitor screen. If the message
Check/calibr. adapter
appears,
perform adapter calibration.
5.
Position the adapter and sensor in upright position in the patient breathing circuit
as close to the patient as possible.
Calibrating the CAPNOSTAT Mainstream sensor
Sensor calibration should be performed whenever the message
Calibrate sensor
appears.
1.
Place the sensor away from all sources of CO
2
, including the patient’s exhaled
breath, your exhaled breath, and ventilator exhaust valves.
2.
With the cable connected to the module, select a gas related parameter window.
3.
Select the
Calibration
tab.
310
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...