Messages
Message
Location
Explanation
What to do
●
Noise
●
wavef.
Arrhythmia alarm category has
been set to
Off
in the
ECG
, and
the ECG channels have not been
available for analysis for the
last 20 seconds or the internal
HR calculation has not been
updated for the last 30 seconds
due to excessive noise which
compromises the accuracy of
detecting events.
●
Remove the source of
excessive noise if possible.
●
Noisy ECG
●
al. area
Arrhythmia alarm category has
been set to
Off
through
ECG
>
Arrhythmia
>
Lethal Alarms
,
and the ECG channels have
not been available for analysis
for the last 20 seconds or the
internal HR calculation has
not been updated for the last
30 seconds due to excessive
noise which compromises the
accuracy of detecting events.
●
Check and remove sources of
excessive noise.
●
Pause
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
●
Please wait -
collecting waveform
●
report
There is not enough ECG sample
data to run the algorithm.
●
Wait for about 10 seconds.
●
QT high
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
QTc high
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
R on T
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
●
Remove one ECG
module
●
al. area
There are two ECG modules in
the system.
●
Remove one ECG module.
●
Single PVC
●
al. area,
wavef.
Physiological alarm.
●
Check the patient status.
●
ST Ant high
/
ST Ant
low
●
ST Inf high
/
ST Inf low
●
ST Lat high
/
ST Lat
low
●
al. area,
param.
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
ST XXX high
/
ST XXX
low
where XXX = ECG lead
label
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
466
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...