Printing
2.
Select the
Arrhythmia
tab.
3.
Select
Print on Alarm
for the arrhythmias you would like to print.
NOTE
For arrhythmia alarm waveform printing, the printing will
continue until 20 seconds has passed from the clearance of
the last active arrhythmia alarm (e.g., 10 seconds saved data,
arrhythmia alarm du 20 seconds data).
Printing waveforms for other than arrhythmia alarms
The following are the default settings for other than arrhythmia alarm waveform
printouts:
●
The print delay is 10 seconds.
●
The print duration is 30 seconds.
●
The waveform speed is 25mm/s.
Other than arrhythmia alarms that print are the high/low alarms for the following: HR,
Art/ABP/Fem/UAC sys/dia/mean, SpO
2
, and ST.
1.
Select
Monitor Setup
>
Printing
.
2.
Select the
Waveforms
tab.
3.
Choose a value from the
Print on Alarm
list:
●
No
: No alarm waveforms print during an alarm condition.
●
High
: Alarm waveforms print during high priority alarm conditions only.
●
All
: Alarm waveforms print during any alarm condition.
Setting the print delay
NOTE
Alarm waveforms (arrhythmia and non-arrhythmia) start
by printing 10 seconds of the most recently saved data,
regardless of the
Delay
setting.
1.
Select
Monitor Setup
>
Printing
.
2.
Select the
Waveforms
tab.
3.
Select a value from the
Delay
list:
a.
0 s
: Manual waveform printing starts with real time data.
b.
10 s
: Manual waveform printing starts with 10 seconds of the most recently
saved data first. After that, the real time data begins to print.
Setting the print duration
NOTE
Non-arrhythmia alarms print 30 seconds of waveforms
regardless of the
Print Length
setting.
NOTE
An arrhythmia alarm waveform continues to print up to 20
seconds after the arrhythmia resolves or a new arrhythmia
has been identified. The length of the event shall determine
the length of the printout.
1.
Select
Monitor Setup
>
Printing
.
2.
Select the
Waveforms
tab.
2062971-001
CARESCAPE Modular Monitors
437
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...