Messages
Message
Location
Explanation
What to do
●
Printing...
●
al. area
Printer: Printing is occurring.
Recorder: Manual printing is
initiated for
Print Waveforms
,
ALL ECG
,
PA Waveform
, or
Catheter Insertion
.
●
Wait for the printing to finish.
●
Recorder cover open
●
al. area
The recorder cover is open.
●
Close the recorder cover.
●
Recorder input
voltage high
/
Recorder input
voltage low
●
al. area
There are problems with the
recorder input voltage.
●
Contact authorized service
personnel.
●
Recorder out of paper
●
al. area
The recorder is out of paper or
the recorder cover is open.
●
Replace recorder paper.
●
Close the recorder cover.
●
Recorder system
error
●
al. area
The local recorder is not working.
●
Reset the local recorder by
turning the monitor’s power
off and on again. If this does
not help, contact authorized
service personnel.
●
Recorder thermal
array overheat
●
al. area
There are problems with the
recorder temperature.
●
Try stopping the recording as
it may help.
●
If the problem persists,
contact authorized service
personnel.
●
Replace battery A
●
Replace battery B
●
al. area
Battery A or B is not working
properly.
●
Replace the battery.
●
Replace monitor
battery
●
al. area
Monitor battery is not working
properly.
●
Replace the battery.
●
Saving
●
al. area
The recorder is unavailable
while printing manual or alarm
waveform recording, and the
recording is saved for later
printing.
No recording location has been
selected.
●
Check the recorder.
●
Select a recording location.
●
Saving to network
●
al. area
Connection to the S/5 Network
was established and the monitor
is saving patient data collected
while disconnected from the
network.
●
No action required.
●
Service Monitor -
and specific error
indication
●
al. area
Technical fault in the monitor.
●
Contact authorized service
personnel.
●
Service the PDM
—
and specific error
indication
●
al. area
Technical fault in the PDM.
●
Contact authorized service
personnel.
●
Setting activation
after next case end
●
al. area
There is a pending setting
activation that will take place
after you end the case.
●
If necessary, cancel the
activation.
2062971-001
CARESCAPE Modular Monitors
511
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...