Messages
Message
Location
Explanation
What to do
●
Replace the module.
●
If the problem persists,
contact authorized service
personnel.
●
ABP disconnect
●
al. area
Invasive pressure line is
disconnected.
●
Check the patient status.
●
Check connections.
●
If pressure drops because of
zeroing, perform the zeroing
process.
●
ABP sys high
/
ABP sys
low
●
ABP mean high
/
ABP mean low
●
ABP dia high
/
ABP dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Art disconnect
●
al. area
Invasive pressure line is
disconnected.
●
Check the patient status.
●
Check connections.
●
If pressure drops because of
zeroing, perform the zeroing
process.
●
Art sys high
/
Art sys
low
●
Art mean high
/
Art mean low
●
Art dia high
/
Art dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Artifact
●
param.
If Smart BP is enabled, this
is normal behavior when
zeroing, flushing, or sampling is
performed.
If Smart BP is not enabled, this
message indicates that the
measurement is disturbed by
artifact.
●
Check the patient status.
●
Check cable contact.
●
Minimize tubing length.
●
Rezero the transducer.
●
Calibrated
●
param.
Channel calibrated successfully.
●
Wait until the message
disappears before starting a
measurement.
●
Calibrating
●
param.
Calibration of a channel is in
progress.
●
No action required.
●
Calibration failed
●
param.
Unsuccessful calibration.
●
Check the connections and
recalibrate.
●
CPP high
/
CPP low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
476
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...