Messages
Message
Location
Explanation
What to do
●
Artifact
●
param.
The sample line length is not
correct.
●
Check that the sample line
length is 2 meters.
●
Change sample line if
necessary.
●
Bypass flow high
●
param.
The module is unable to
synchronize flow and CO
2
due to
bypass flow.
●
Add a 5 ml spacer to the
patient circuit between the
Y-piece and D-lite.
●
Shorten the expiration time to
avoid expiratory flow pause.
●
No VO2, FiN2O high
●
param.
Module has detected N
2
O.
●
Gas exchange cannot
be measured if N
2
O is
used. If you wish to
measure gas exchange,
use another anesthetic. If the
measurement is not used,
you may consider removing
gas exchange numbers from
the screen.
●
No VO2, FiO2 >85%
●
param.
Measured FiO
2
is more than
85%.
●
Gas exchange cannot be
measured if FiO
2
is more than
85%. If the patient’s oxygen
values are consistently high,
you may consider removing
gas exchange numbers from
the screen.
●
Out of range
●
param.
VO
2
or VCO
2
is below 0 ml/min
or above 999 ml/min.
●
Check that the gas sampling
line and spirometry lines are
correctly connected to the
patient airway and to the gas
module.
●
Check that the correct sensor
type (D-lite/Pedi Lite) has
been selected from the
monitor menu.
●
Service gas module
●
Service gas module
- and specific error
indication
●
al.area,
param.,
wavef.
The measuring sensor is
inoperative or the temperature
in the module has increased.
●
Contact authorized service
personnel.
Messages related to Entropy measurement
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
496
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...