Snapshots and events
need more memory than the previous ones. Therefore, more than one snapshot
may have to be erased.
To erase snapshots, you must discharge the patient/end a case. Note that this will not
erase just the snapshots but also all other patient and trend data from the monitor
and a connected PDM or TRAM, and will also return monitor settings to their defaults.
1.
Select the patient information area on screen, or select
Data & Pages
>
Start
/ Reset Case
(OR and PACU software packages) or
Admit/Discharge
(other
software packages).
2.
Select
Patient
>
Discharge Patient
.
3.
Select
Yes
.
Snapshots and alarm history
CAUTION
— MEASURING DATA STORED IN ALARM HISTORY — Waveform
data is stored in the alarm history using compression
technology that may not allow perfect reconstruction of
the waveform data when subsequently viewed. Although
differences occur relatively frequently and are usually
very minor, users are urged to verify diagnostic waveform
measurements with the waveform data from realtime graph
strips.
When a PDM or telemetry transmitter is connected to the monitor, the saved alarm
history and ST history will be transferred to the monitor. One snapshot is created
for each alarm history.
The PDM or a telemetry transmitter alarm history has 10 seconds of waveform data
from two or three ECG leads and the first arterial invasive pressure line. The snapshot
will display waveforms only if the snapshot field configuration includes the same
ECG waveforms and/or arterial invasive pressure channel. Otherwise, no waveform
data is displayed.
A snapshot is sent to the CARESCAPE Network if it contains one of the following
waveforms: ECG lead I, II, III, or Va, or any arterial invasive pressure.
Snapshot transfer to PDM
Telemetry transmitters store snapshots independently. They are stored on the PDM
when they are transferred by the monitor. Only snapshots created by arrhythmia
alarms are transferred. Manual snapshots and snapshots triggered by SpO
2
high/low
and arterial IP high/low alarms are not stored to the PDM.
ST snapshots
Creating ST snapshots manually
An ST snapshot displays QRS complexes.
1.
Select
Monitor Setup
>
Parameter Setup
.
2.
Select
ECG
>
ST
.
3.
Select
Realtime View
.
428
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...