ECG
WARNING
This device uses a computerized 12 lead ECG analysis
program, which can be used as a tool in generating ECG
records that provide ECG measurements and interpretative
statements from the ECG recordings. The interpretive
statements are only significant when used in conjunction
with clinical findings. All ECG records should be overread by
a qualified physician. To ensure accuracy, use only the ECG
records for physician interpretation.
WARNING
When transitioning from a 10-lead cable to a 5-lead cable,
select the
Update Lead Set
option to clear the
Lead off
message from the display.
WARNING
Disconnected electrodes or loose electrode connections can
lead to missed critical severity ECG alarms. If the monitor
reports
Leads off
after selecting
Update Lead Set
option,
always check the electrode connections to the patient.
WARNING
— CONDUCTIVE CONNECTIONS — Extreme care must be
exercised when applying medical electrical equipment. Many
parts of the human/machine circuit are conductive, such
as the patient, connectors, electrodes, transducers. It is
very important that these conductive parts do not come
into contact with other grounded, conductive parts when
connected to the isolated patient input of the device. Such
contact would bridge the patient’s isolation and cancel the
protection provided by the isolated input.
WARNING
— DELAYED ASYSTOLE ALARM — The pulsatile heart rate may
have a slower response time than the electrical heart rate
where there is a low perfusion patient condition. When using
the IntelliRate feature in this situation, the monitor may delay
calling an ASYSTOLE patient alarm. The user may elect to turn
the IntelliRate feature off for patients at risk of these events,
otherwise patient treatment may be delayed. Such patients
should always be kept under close observation.
WARNING
— NOISY ECG alarm — The
Noisy ECG
alarm indicates that the
system is no longer monitoring ECG and there may be no
HR
high
,
HR low
,
Tachy
or
Brady
alarms. If you adjust the alarm
priority level lower than the default value, keep the patient
under close surveillance.
WARNING
— INACCURATE HEART RATE INDICATION — PDM — The
electrical and pulsatile heart rate values provided by the
various monitored parameters (ECG, SpO
2
, blood pressures)
may differ markedly. These differences may be due to
underlying physiologic conditions (e.g., electromechanical
dissociation, pulseless electrical activity, non-perfusing
rhythms) or to inaccuracies in the heart rate values caused
by artifact or poor signal quality. The user may elect to turn
the IntelliRate feature off for patients at risk of these events,
otherwise patient treatment may be delayed. Such patients
should always be kept under close observation.
148
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...