Safety
WARNING
If an error message appears during operation, it is the licensed
medical practitioner’s responsibility to decide whether the
unit is still suitable for patient monitoring. As a general rule,
monitoring should only continue in extremely urgent cases
and under the direct supervision of a licensed healthcare
practitioner. The unit must be repaired before being used
again on a patient. If an error message appears after
power-up, the unit must be repaired before being used on
a patient.
WARNING
Ensure that modules are securely latched.
WARNING
The parameter modules are not able to withstand unpacked
drops from a height of 1 m without damage. If a module is
dropped, please service it before taking it back into use.
WARNING
If the frame or monitor is dropped, please service it before
taking it back into use.
Site requirement warnings
WARNING
— BEFORE INSTALLATION — Compatibility is critical to safe and
effective use of this device. Please contact your local sales or
service representative prior to installation to verify equipment
compatibility.
System caution safety messages
The following caution safety messages apply to this monitoring system.
Loss of data
CAUTION
LOSS OF DATA — Should the monitor at any time temporarily
lose patient data, the potential exists that active monitoring
is not being done. Close patient observation or alternate
monitoring devices should be used until monitor function
is restored. If the monitor does not automatically resume
operation within 60 seconds, power cycle the monitor
using the power on/off switch or on/standby button. Once
monitoring is restored, you should verify correct monitoring
state and alarm function.
Electrical caution
CAUTION
— POWER REQUIREMENTS — Before connecting the device
to the power line, check that the voltage and frequency
ratings of the power line are the same as those indicated
on the device’s label. If this is not the case, do not connect
the system to the power line until you adjust the device to
match the power source. In U.S.A., if the installation of this
equipment will use 240V rather than 120V, the source must be
a center-tapped, 240V, single-phase circuit. This equipment is
suitable for connection to public mains as defined in CISPR 11.
46
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...