Messages
Message
Location
Explanation
What to do
●
CVP sys high
/
CVP sys
low
●
CVP mean high
/
CVP mean low
●
CVP dia high
/
CVP dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Disconnected
●
param.
Pressure is below physiological
detection threshold.
●
Check the patient status.
●
Check the cable and
connections.
●
If pressure drops because of
zeroing, perform the zeroing
process.
●
Fem disconnect
●
al. area
Invasive pressure line is
disconnected.
●
Check the patient status.
●
Check connections.
●
If pressure drops because of
zeroing, perform the zeroing
process.
●
Fem sys high
/
Fem sys low
●
Fem mean high
/
Fem mean low
●
Fem dia high
/
Fem dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
FemV sys high
/
FemV sys low
●
FemV mean high
/
FemV mean low
●
FemV dia high
/
FemV dia low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
ICP sys high
/
ICP sys
low
●
ICP mean high
/
ICP mean low
●
ICP dia high
/
ICP dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Identical IP modules
●
Identical IP2 modules
●
al. area
There are two or more IP
modules in the system.
●
Remove all but one IP module.
●
Identical IP3 modules
to
●
Identical IP8 modules
●
al. area
There are two or more IP
modules in the system.
●
Remove all but one IP module.
●
Identical IP RAC
modules
●
al. area
There are two or more IP RAC
modules in identical slots in the
system.
●
Remove all but one IP RAC
module.
2062971-001
CARESCAPE Modular Monitors
477
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...