4.
Ethernet: For connecting the MC and IX networks. The MC Network establishes
communication and allows patient data to be sent to an optional CIC Pro Clinical
Information Center or CARESCAPE Central Station. The IX Network provides access
for example to the MUSE server, Citrix server, and IX printers.
5.
Tram-Net and ePort: For connecting the PDM, E-module Frame, and Tram-Rac
housing.
6.
RS232: For connecting the touchscreen display.
7.
DVI-1 supports a digital display and cloned analog display. DVI-2 supports only
one digital display. DVI-3 (optional third video) supports a digital display and a
cloned analog display (iPanel application only).
8.
Ground plug: Provides equipotentiality.
B850 module frames F7 and F5
WARNING
Do not touch the electrical connector located within the
module housing or frame.
WARNING
— B850: ELECTRIC SHOCK — Do not use the F7 Frame for
standalone use. Ventilation holes on the F7 E-module Frame
will be covered only if installed within an Aisys, Avance, or
Aespire anesthesia machine.
1.
Defibrillator synchronization
F5 and F7 E-module Frames provide an interface between the monitor and E-modules.
Frames allow additional parameters to be monitored.
The F5 Frame has five module slots that support all E-module acquisition modules. It
supports both PDM and PSM with a slide mount.
The F7 Frame has seven module slots, but it does not have a slide mount for the PDM
or PSM. The PSM module may be interfaced to the F5 or F7 Frame with a cable when
the Module Bus Adapter or Interface Module for PSM is used.
2062971-001
CARESCAPE Modular Monitors
57
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...