Neuromuscular transmission
Problem
Solution
Difficulty in getting the response when locating the
nerve for plexus stimulation
●
You may try using the local muscle response as an
indication of current in the needle. If there is no
response, the needle may be broken.
●
Change the needle if necessary.
How can I verify the measurement reliability?
●
The reliability of the response can be estimated by
observing the bar graph or the NMT trend.
●
The bars of the bar graph should be in a smoothly
descending order from left to right, and the NMT
trend should indicate a fairly stable T1%. If this
is not the case, the latest response is unreliable.
Relaxation level does not usually decrease
significantly in one minute even with short-acting
relaxants.
●
To verify your observations, start a new
measurement manually right after the completion
of the previous measurement. Sometimes
the selected cycle time has been so long that
the relaxation level has changed considerably
between the measurements.
What can I do if the monitor does not find the
supramaximal stimulation current?
●
If the monitor is unable to find the supramaximal
stimulation current, check the electrode
placement. The nerve may be outside the dense
current flow or both the ulnar and median nerves
may be stimulated; progressively more muscle
activity is detected as the increasing stimulation
current activates new motor units. Consult the
figure below when placing the electrodes:
1. adductor pollicis
2. abductor digiti minimi (hypothenar)
3. flexor pollicis brevis (thenar)
4. ulnar medial
5. epicondyle
●
Also, supramaximal current may not be found if
the patient has anatomical and/or physiological
anomalies.
2062971-001
CARESCAPE Modular Monitors
371
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...