Messages
Message
Location
Explanation
What to do
●
Error 1: PMC update
failed. Turn the
monitor off and then
on again.
●
start-up
screen
Unsuccessful PMC update.
●
Turn the monitor off and then
on again.
●
If the problem persists,
contact authorized service
personnel.
●
Error 2: PMC update
requires a mains
supply. Plug in the
power cord and turn
the monitor off and
then on again.
●
start-up
screen
Unsuccessful PMC update.
●
Plug in the power cord and
turn the monitor off and then
on again.
●
If the problem persists,
contact authorized service
personnel.
●
External alarm light
disconnect. Check
USB connection.
●
al.area
There is a communication error.
●
Wait and see if the message
disappears.
●
Try selecting audio pause
and see if the message
disappears.
●
If the problem persists,
contact authorized service
personnel.
●
Identical IP address
noticed
●
al. area
Two or more monitors have the
same IP address.
●
Contact authorized service
personnel.
●
Identical unit & bed
name noticed
●
al. area
Two or more monitors have the
same unit and bed name.
●
Contact authorized service
personnel.
●
Incompatible Module
●
al. area
The module is not compatible.
●
Replace with compatible
module. See the CARESCAPE
Modular Monitors
Supplemental Information
Manual.
●
If the problem persists,
contact authorized service
personnel.
●
Incompatible
TramRAC
●
al. area
Incompatible Tram-Rac
connected.
●
Replace with compatible
Tram-Rac module. See the
CARESCAPE Modular Monitors
Supplemental Information
Manual.
●
If the problem persists,
contact authorized service
personnel.
●
Incorrect barcode
value
●
al. area
The barcode string differs from
the defined values.
●
Contact authorized service
personnel.
●
Information Center
busy
●
al. area
S/5 Network has too many
bed-to-bed connections.
●
Try again later.
●
Information Center
not responding
●
al. area
S/5 Network is busy and the
central station is not responding.
●
Try again later.
●
Invalid barcode
configuration
●
al. area
The barcode configuration is not
correct.
●
Contact authorized service
personnel.
508
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...