K465i Operations and Maintenance
P/N 1212831 Rev C
Page 5-43
Copyright
©
2012
Veeco Instruments, Inc. Confidential
All Rights Reserved
16. Vacuum the shutter by using a
dry
cleanroom wipe to wipe the shutter, and hold the vacuum close
by to catch the debris.
Figure 5-39. Vacuuming of the Shutter
17. Using a long q-tip (cotton swab on a plastic stick), clean the RT/DRT viewport from the bottom of the
flow flange.
18. Vacuum the bottom of the flow flange, ensuring that the vacuum nozzle does not come in direct
contact with the flow flange (keep ~ ¼ inch distance between vacuum nozzle and flange).